黄瓜污视频导航,黄瓜视频黄片,黄瓜视频色版,黄瓜视频APP免费观看下载安装

Product Name: HT-TD430 Humidity Transmitter
器特點
Instrument Features

a. Compact and aesthetically pleasing in appearance, with a unique structural design;

b. High-resolution LCD display provides clear and beautiful visuals;

c. It could offer a cost-effective solution for the detection of trace moisture in high-end glove boxes;

d. Resistant to a small number of corrosive gases;

e. It has sensor over-limit protection and automatic recovery functions;

f. Menus are locked to prevent accidental changes to instrument parameters and affect its performance;

g. It is available for installation using KF40 flanges or threaded connections for in-situ measurement, or using a test chamber for ventilation measurement;

h. The circuit employs digital processing technology, isolation amplification, and band-pass design, featuring strong anti-interference capabilities;

i. RS485 bidirectional communication function enables direct communication with computers or other digital communication devices.

術參數
Technical Parameters

l Measurement Principle: Electrolyte

l Display Mode: LCD display with high resolution

l Measurement Range: 0~500ppmV H2O (range customizable)

l Measurement Accuracy:0~10ppm±3%FS,10~500ppm±5%FS

l Repeatability:±2% (when 10ppm)

l Stability:±2%FS/7d

l Resolution: 0.01PPm

l Response Time 63%: T63≤3min

l Output Port: 4-20mA (load resistance<1KΩ)

l Communication Mode: RS485

l Power Supply: DC24V±10%, 0.2A

l Storage Environment Humidity: <90%RH, non-condensing (When the storage and transportation environment temperature is lower than 5℃, the transmitter should be placed in a normal working temperature environment for more than 4 hours before use.)

l Operation Environment Humidity: <100%RH, non-condensing

l Sample Mode: Free diffusion (other method is customizable)

l Sample Gas Temperature: -10℃~+50℃

l Coating Cycle: 6 months (The higher the concentration measured, the shorter the coating cycle.)

l Calibration Cycle: 1 year (recommended)

l Sensor Service Life: >24 months (under normal conditions of use)

l Gas Path Connection: NPT 1/8 internal thread for calibration cover

l Installation Method: KF40 flange installation

用場合
Application Scenarioses

It is used in glove boxes, research and production of lithium-ion batteries, inert atmosphere welding, scientific laboratories, pharmaceuticals, nuclear industry, and chemical applications.

貨須知
Ordering Instructions
Contact Number

028-85044052/053

Top
網站地圖